Andersen Cascade Impactor
6 Stage Viable

Andersen Cascade Impactor 6 Stage TCR Tecora
1
2
3
4
5
6
7
8
Andersen Cascade Impactor 6 Stage Viable
1

1st Stage d50 = 7 μm

2

2nd Stage d50 = 4.7 μm

3

3rd Stage d50 = 3.3 μm

4

4th Stage d50 = 2.1 μm

5

5th Stage d50 = 1.1 μm

6

6th Stage d50 = 0.65 μm

7

Produced with AISI 316

8

Italian FlagMade in Italy

Andersen Cascade Impactor 6 Stage Viable TCR Tecora

In accordance with

EN 14683
ASTM F2101

Andersen Cascade Impactor 6 Stage Viable

Aerodynamic Particle Sizing

The design concept of the Multistage Viable Impactor evolves from the following information:

The human respiratory system tract is an aerodynamic classifying system for airborne particles. A sampling device can be used as a substitute for the respiratory tract as a collector of viable airborne particles, and as such, it should reproduce to a reasonable degree the lung penetration by these particles.

Viable particle samplers must satisfy the following criteria:

  • Particles must be separated from the air and “physically” captured onto a suitable impaction surface (particle collection efficiency);
  • For viable bioaerosols, the viability is critical and shall be maintained during sampling to be representative of actual levels of exposure to the pathogen (biological collection efficiency).

The 6 Stage Multistage Viable Impactor with 400 small round jets per stage meets all the criteria for the efficient collection of airborne viable particles.

Bioaerosol

The study of the microbial content of the air has become increasingly more significant in the past decade as the needs for “contamination free” environments have become more apparent.

Biological aerosols have been defined as biological contaminants occurring as solid or liquid particles in the air. These particles can vary in size from viruses less than 0.1 micron in diameter to fungal spores 100, or more microns in diameter. They may occur as single, unattached organisms or as aggregates.

The 6 stage Viable Multistage Impactor is a multi-orifice, cascade impactor which has been widely used as a standard to measure the concentration and particle size distribution of culturable bacteria and fungi in ambient air as well as other bioaerosols, like viruses.

Main Features

  • Aerodynamic particle sizing;
  • Sampling flow: 28,3 L min-1;
  • Particles are collected directly in the Petri (ø 90 mm) filled with the most appropriate culture medium;
  • Material: AISI316 (Aluminum on request)
  • Height: 202mm;
  • Diameter: 105mm;
  • Weight: 3Kg (AISI) - 1,25Kg (Aluminium);
  • Carrying case;
  • Dimensional Ispection Certificate;

Products Codes

ProductsCodes
6 Stage Viable Cascade Impactor (IMP-6BIO)AC99-120-0002SP
CF qty 18 Pyrex® glass Petri Dishes ø 90 mmAC99-120-0006SP
CF qty 720 Plastic Petri Dishes ø 90 mmAC99-120-0007SP
Constant Flow Sampler Bravo BIOAA99-000-0030SP (Bravo Basic BIO)
AA99-000-0740SP (Bravo X BIO)

TCR Tecora has developed a complete BFE Test Kit in accordance with EN 14683, ASTM F2100, ASTM F2101 and EN ISO 9237.